Bạn đang xem bản rút gọn của tài liệu. Xem và tải ngay bản đầy đủ của tài liệu tại đây (2.48 MB, 40 trang )
93
Membranes
Further characterisation is reported in the organics characterisation section below. An overview over
some characteristics is also shown in Chapter 2. The organics were prepared as 100 mgL-l organic
carbon stock solutions by mixing the dry powder with MilliQ water without increasing the pH. The
solutions were stored at 4°C in the dark. The amount of powder required for 100 mL stock solution
was 18.4 mg, 18.6 mg and 200 mg for HA, FA and NOM respectively. This reflects the carbon content
of the organics.
Hematite was selected as a model colloid in this study due to its well understood aggregation behaviour,
the monodisperse, spherical nature of the colloids and the fact that the synthesis of colloids of various
primary particle sizes (40 to 500 nm) is possible. WGLlle silica and clays may be more abundant in
surface waters, hematite appears to be a good compromise between real systems and a simple model
compound.
The synthesis of monodispersed, spherical hematite colloids of four primary particle sizes is described
in detail in Appendix 3. The main properties of these colloids are also given in Appendx 3.
Commercially available flat sheet membranes were selected. The primary selection criterium was that
the membrane be made of a hydrophlic material, which adsorbs less organics than more hydrophobic
polymers. For comparison, the membranes used are listed in Table 4.2 with their pore size or molecular
weight cut-off (MYVCO) as specified by the manufacturer.
Table 4.2 Characteristics OfMF, UF and NF membranes used in experiments.
Process
Mdhpore
Type
GVWP
GVHP
UF
hblhpore
PLHK
PLTK
PLGC
PLCC
PLBC
PWC
Fluid
CA-UF
Systems
NF
TFC-SR
TFC-S
TFC-ULP
Copyright © 2001 by Andrea I. Schafer
Typical
Specifications
Pure Water
Surface Charge
Operating
Pressure
[bar1
MF
Supplier
Pore Size [pm]
Molecular Weight
cut-off p a l
Flux
Fm-2h-l]
at p H 8
94
MATERIALS AND METHODS
This characterisation is relatively vague, as different methods are used by each manufacturer (Readman
(1991), Thorsen e t al. (1997)). As a more comparable parameter, the pure water fluxes as determined in
the experiments are also given, as well as the membrane zeta potential at pH 8. A new membrane was
used for each experiment (except for fractionation experiments).
The results of surface charge measurements of the membranes as a function of pH, pure water fluxes
and electronmicrographs are shown in the MF, UF, and N F chapters, respectively.
4.4.1
Microfdtration Membranes
Two microfiltration membranes @hllipore, hydrophilic ( G W ) and hydrophobic (GVHP)) with
nominal pore sizes of 0.22 pm were used. The hydrophlic membrane is a modified hydrophobic
membrane. The hydrophilic membrane was chosen for most experiments because hydrophlic
membranes have a reduced adsorption capacity towards hydrophobic organics (Jucker and
Clark(1994)).The membrane material is a modified polyvinylidene fluoride (PVDF).
The hydrophobic membrane was soaked in a 50% ethanol solution for 10 minutes to wet the pores and
then rinsed with MilliQ water. All membranes were soaked in warm MdliQ water for 30 minutes prior
to use to remove any organic contamination.
4.4.2
Ultrafiltration Membranes
Ultrafiltration was used for fouling, rejection, and fractionation experiments. The fractionation
experiments require membranes with very low adsorption characteristics to reduce loss of organics on
the membranes. It was thus necessary to find low fouling membranes, whch are available in a range of
membrane molecular weight cut-offs QWXCO). The fillipore "PL series" fulfil the low adsorption
condtion and they are available in seven MWCOs in the range from 1 kDa to 300 kDa. The
fractionation membranes selected were the PLAC, PLBC, PLCC, PLGC, PLTIC, and PLHIC with
MWCOs of 1, 3, 5, 10, 30, and 100 kDa, respectively. Fouling and rejection experiments were carried
out with the 10 and 100 kDa membranes.
These regenerated cellulose membranes on a non-woven polypropylene substrate are described by the
manufacturer as low protein-binding and hydrophlic. The MWCO (as described in Table 4.2) is
determined by a range of Dextran markers. A MWCO of 10 kDa means that 90°/o of markers with a
molecular weight greater than 10 kDa were retained.
Prior to use, the membranes were soaked in 0.1 M NaOH for 30 minutes and flushed with 3.4 L of
MilliQ water in order to remove the glycerin preservative, whch can strongly interfere with UV and
DOC analysis. Alternatively, flushng the membrane with 1L fiUiQ also removed the glycerin
sufficiently.
4.4.3
Nanofdtration Membranes
Nanofiltration membranes were received from Fluid Systems in San Diego, USA (now Koch
Membrane Systems). T h n film composite membranes were chosen due to their low fouling
characteristics compared to polysulphone membranes used in other studes. The CA-UF membrane is,
as the name suggests, classed as a UF membrane and the material is cellulose acetate. However, it is
treated as a NF membrane here as it is often used for similar applications according to the
manufacturer, and also because it exhibits some salt rejection. Membrane characteristics as given from
Copyright © 2001 by Andrea I. Schafer
95
Membranes
the supplier are summarised in Table 4.3. The cut-off was specified to be about 5 kDa and the material
is non-ionogenic. The active layer of this membrane is about 150 nm. CA membranes have generally a
50% lower flux than TFC membranes, but are cheaper.
The TFC membranes are chemically modified to render the membranes more hydrophilic, but more
details were not available. All three membranes have different additives and post-treatments in the
manufacturing process. The manufacturer estimates the thckness of the active layer of the TFC
membranes to be 150 to 200 nm. For the TFC-SR membrane a dfferent monomer was used compared
to the other TFC membranes. \%le
the TFC-S and TFC-ULP membranes are made from
metaphenylene diarnine with acid chloride (a benzene ring with two to three carboxglic acid groups),
the TFC-SR membrane is fabricated from a mixture of cyclo-aliphatic amine with acid chloride. This
means that the TFC-S and TFC-ULP have both positive and negative functional groups, whereas the
TFC-SR membrane has negative functional groups only. Marker tests with 1% lactose (180 Da)
solutions at pH 6-7 showed a rejection of 94.4% and 90.6% for the TFC-SR and TFC-S membranes,
respectively. Rejection of the membrane is expected to be higher ('I'akigawa (1999)).
Table 4.3 Membrane Infornationfrom Flziid Systems Corporation (now Kocb Membrane Systems), San Diego.
TFC-S
Material
Test
Condttions
TFC-SR
TFC-ULP
CA-UF
TFC P-1 on PS base
Cellulose Diacetate
1 g/L NaCl
2 g/L NaC1
tap water 3.5 bar
25°C pH 7.5 7 bar
TFC Polyamide (PA\) TFC proprietary P-\ on PS
base, coated with PT';\
on Polpsulfone (PS)
base
(dye to check for damage)
1 g/L NaC1,
Flux
2.5 g/L AlgSO,
2.5 g/L AlgSO,
25OC pH 7.5 5.6 bar
25OC pH 7.5 5.6 bar
14.7 L/m'h
14.7 L/m2h
14.7 L/m2h
16.5 L/m2h
98.5% hardness,
98.5% C
1
Not specified
N F or softentng of
municipal water at
dtralow pressure; up
to 45°C
nanofiltration or softening
of municipal water at
ultralow pressure; up to
1 ppm C12; up to 45 "C
Industrial or municipal
water
Surface water at
moderate pressure if
chlorination desired
(up to 1 ppm C12)
5.6 bar
5.6 bar
3.5-12.25 bar
p H range
4-11
Rejection
95°/a hardness,
4-6
85% C1
Design
Application
Design
Pressure
(560 kPa)
Storage
0.5% sodmm meta
bisulfite, ALilliQ after
ultralow pressure
Medium
3.5 bar
(560 kPa)
unknown
ALilliQ after wash
SUiQ after wash
wash with warm AlilliQ to
remove PT',I coating
soak in XLdliQ
wash
Pretreatmen t
wash wlth SIdhQ
Copyright © 2001 by Andrea I. Schafer
wash with ALtUIQ
96
MATERIALS AND METHODS
All membranes were stored in a refrigerator (4 K ) in plastic bags in the medium in which they arrived,
and sealed. A few membranes of each type were cut out, pretreated and then placed in a Petri dish in
the refrigerator for use in experiments.
Stirred cell systems were selected for the experimental work for a number of reasons; (i) volumes are
small whch is required for the use of IHSS reference material, (ii) membrane samples are small which
allows the use of a new membrane for each experiment, (iii) the solution chemistry can be precisely
controlled, (iv) experiments are relatively short and thus the investigation of a great number of
parameters is possible, and (v) the concentration in the cell represents the concentration in a crossflow
module (recovery about 70%). A comparison of mass transfer values was demonstrated in the case of
NF in Chapter 7. Drawings of the filtration equipment are shown in Appendix 2. A hydrodynamic
analysis is also shown in Appendix 2.
4.5.1
Microfiltration Equipment
All experiments were carried out in a magnetically stirred batch cell (volume of 110 mL, membrane area
15.2 . 10-4m" at a pressure of 100 kPa (if not otherwise indcated), pressurised with nitrogen gas. A
reservoir of 1.5 L volume was connected to the stirred cell. A photo of a Perspex stirred cell with
reservoir, manufactured in the university workshop, is shown in Figure 4.1.
Figure 4.1 Perqex stirred cell w t reservoir.
ih
All stirred experiments were stirred at 270 rpm (measured with a Philips PR 9115/00 stroboscope). A
balance and stop watch were used to measure permeate volume. Experiments were conducted at a
temperature of 25
+ 1 OC.
Copyright © 2001 by Andrea I. Schafer
Filtration Equipment
4.5.2
Ultrafiltration Equipment
The same system as described in the MF section and shown in Figure 4.1 was used for all rejection,
fouling, and fractionation ultrafiltration experiments. The balance was connected to a PC for flux data
collection.
4.5.3
Nanofiltration Equipment
Nanofiltration experiments were carried out in a stainless steel stirred cell with an Amicon magnetic
stirrer on a magnetic heater plate (Industrial Equipment & Control, Australia). The calibration is
shown in Figure 4.2.
The volume of the cell was 189 mL, the inner dameter 56.6 mm (resulting in a membrane surface area
of 21.2 10-"%
The stirrer speed could be varied from about 200 to 2000 rpm, with a setting of 400
rpm used routinely. The stirrer speed was measured using a Phlips PR 9115/00 stroboscope. One
side of the stirrer bar was labelled to avoid measuring of half rotations.
Figure 4.2 Calibration of magnetic stirrer table.
Figure 4.3 Stainless steel stirred cell set-zp.
Copyright © 2001 by Andrea I. Schafer
98
MATERIALS AND METHODS
The stirred cell was pressurised with instrument grade air. An- was used (rather than N 2), to provide
C02 for the carbonate buffer. pH changes due to the high pressure air were estimated to be less
significant than with N2 (see Appendix 5 for details). A photo of the set-up is shown in
Figure 4.3
and a schematic in Figure 4.4.
The cell was equipped with a pressure gauge mounted in the stainless steel line after the air cylinder, a
stainless steel reservoir with a volume of 2 L, a pressure release valve, a fluid inlet and outlet
connection, a pressure safety valve, and a refill opening on top of the reservoir. O n top of the stirred
cell, a fluid inlet connection, a pressure release valve and a temperature probe fitting were mounted.
The temperature was measured with a PT 100 probe, connected to a Kane-May ISM 330 indicator.
To control the temperature inside the cell, it was placed in a 2 L plastic beaker, through whch tap
water was circulated continuouslp. The temperature was kept constant (unless otherwise indicated) at
20 "C k 1 ()C. Permeate flux was measured by weight with a Mettler-Toledo PR 2002 (0.1 to 2100g)
balance, whch was connected to a PC equipped with Mettler-Toledo BalanceLink software.
Figure 4.4 Stainless steel stirred cell set-up. A : stirred
cell: z~olme185 mL; B: magnetic stirrer (Amicon, dtiven
bJy magnetic stirrer table); C: membrane; D: stainless steel
porons support; E: reseruoir uolnme 2000 mL, F:
pressurired instrtlment air inlet, G: feed inlet, presswe
release and safe9 valves; H: permeate outlet (to balance
and PC).
4.6.1
p H Value
A Beckrnan glass electrode (Ag/AgCl) was used for solution preparations and no contamination was
observed. The electrode was only used in samples after DOC analysis and was cleaned prior to use for
pH adjustment.
4.6.2
Conductivity
Conductivity was measured using a Lutron CD-4303 portable instrument.
4.6.3
Inductively Coupled Plasma Atomic Emission Spectroscopy (ICP-AES)
A Perkin Elmer Optima 3000 Spectrometer was used to determine the cation content of solutions.
Samples and multielement standards (0, 1, 10 and 100 mgL-l) were diluted with 5% nitric acid. All vials
used were cleaned with 1 M sulphuric acid. Detection limits are 3, 5,0.1, 5, and 70 pgL-' for Fe, Al, Ca,
Na, and I<, respectively.
Copyright © 2001 by Andrea I. Schafer
Organics Characterisation
99
The particle sol and filtration samples were diluted 1:l with HC1 (36'Yo) and heated (in a closed sample
vial) to dissolve the colloidal hematite. These samples were then analysed directly.
4.6.4
Ion Chromatography (IC)
IC was used for chloride determination for N F rejection experiments. Anions could not be analysed
as
using ICY humic substances interfere with the analysis (Hoffmann et al. (1986)). A Millipore Waters
Model 590 instrument was used with a Model 430 Conductivity detector. The eluent used was 0.68 gL -1
boric acid (H;BO3), 0.235 gL-' gluconic acid anhydride (C6H1006) and 0.3 gL-1 lithum hydroxide
@OH - 6 HzO).
4.7.1
Dissolved Organic Carbon (DOC)
Dissolved organic carbon was analysed using a Skalar 12 instrument. The method is based on UVpersulphate oxidation and described in detail in Appendix 4. The DOC of every sample was measured
as a routine analysis.
For samples containing colloids, aggregates or flocs the measured value is total organic carbon (TOC).
None of the samples were filtered as this would lead to loss of organics.
4.7.2
UV/VIS Spectroscopy
A Varian Cary 1E UV/VIS Spectrophotometer was used to evaluate the method and for further
standard analysis. Spectra of UV/VIS in the range from 190 to 500nm were obtained and correlations
established with DOC analysis. The method is further described and evaluated in Appenchx 4. UV/VIS
was also a routine analysis and the wavelength was used in rejection calculations.
At low wavelength (190 nm region), absorption by inorganics is observed. This is strong in the case of
unpurified Mooney Mooney NOM and absent in the purified IHSS samples. The ion content of all
samples is shown in section 4.7.6.
The W / V I S spectrum of NOM is attributed mainly to absorption of light energy by aromatic
compounds and can be broken into a series of transition bands, similar to those published for benzene
(Korshin et al. (1997b)). Three transition bands can be distinguished for each aromatic chromophore in
NOM - the local excitation (LE) band, the benzenoid ( ' 2 ) band, and the electron-transfer (ET) band.
The peaks vary in their height, width, and centre location depenchng on the composition of the NOM
(Kaecbng (1998)). The presence of these various peaks can be recognised in the shoulders on the
spectra as shown in Figure 4.5, however detailed analysis was not considered warranted.
From Figure 4.5, it can be seen that the (probably) soil-derived fidrich HA (purified with a lOOkDa UF
membrane) has the largest UV/VIS absorbance, followed by IHSS and the NOM HA fraction whch
are surprisingly similar. The FA fraction of Mooney Mooney NOM has a higher absorbance than the
unpurified NOM, which can be explained given the NOMs relatively high content of hydrophlic acids
of a very low absorbance. The IHSS FA also has a slightly lower absorbance over the complete
wavelength range.
Copyright © 2001 by Andrea I. Schafer
100
MATERIALS AND METHODS
5 0.15
Figure 4.5 C V Spectra of the organics zmd.
all wavelengths linear with concentration
IHSS FA
NOM
IHSS HA
Aldrich HA (c100 kDa)
NOM Hydrophilic Fraction
NOM HA Fraction
2
4.7.3
200
250
300
350
400
450
500
UVIVIS Wavelength [nm]
Titration
The NOhl sample, which was concentrated as described in Appendix 1, was titrated using a Metrohm
automatic titrator. .The titrator was operated in dynamic titration mode. The samples were acidified
from ambient pH to pH 2.8 with 0.1 M HNO3 and subsequently alkalised with 0.1 M NaOH to pH 10.
It was assumed that at pH 2.8 all acidic functional groups will be saturated, whereas at pH 10 all
carboxylic and half of the phenolic groups were dissociated. The limitations of these assumptions were
discussed in Chapter 2.
The titration vessel was purged with nitrogen to eliminate C 0 2. From the volume and molarity of
added base and the mass of titrated DOC, the content of acidic functional groups can be calculated.
Carboxylic acid content was calculated from the amount of base added until the end-point was reached.
Phenolic acid content was calculated as twice the difference in titrant required to change the pH of the
titrate from 8 to 10, since it was assumed that at pH 10 only half the phenolic groups were Issociated.
A solution of a concentration of 20 mgL-I as DOC NOM were titrated. The error due to the salt
content of NOM is likely to be high.
Table 4.4 describes the a c i d q and size of the three organics used and the average molecular weight as
found in the literature (for IHSS organics and purified Aldrich HA) or as measured (for NOW. The
reported rvlW will be verified later (see section 4.7.7) by analysis.
Table 4.4 Acid$ and average molecdar weight ofthe organics ( y ~ c k e r Clark (1984),'Beckett et al. (1987),
and
y
'Elering and Morel(1988),'ana&red b titration (lee above), 'Clark andhcker (1993), 'Children and Elimelecb
(1996)).
Type of Organic
Acidq [meq.gl]
Carboxylic
liverage Molecular Weight p a l
Phenolic
IHSS FA
3.41 5.45 6.1
1.512.05
IHSS HA
4.01 4.16
2.9' 2.16
Purified Aldnch L A
3.3(j
2.56
> 50 0006
Mooney Mooney NOM
5.14
1.34
< 10004
Copyright © 2001 by Andrea I. Schafer
7502
1100"
15002
12004
101
Organics Characterisation
4.7.4
Elemental Analysis
Elemental analysis of the IHSS reference material was provided by IHSS with purchase of the organic
material. The elemental analysis was performed for IHSS by Huffman Laboratories (Wheat Ridge, CO,
USA). Results are summarised in Table 4.5.
Table 4.5 Elemental analy,is resztlts ofthe organics used.
C
Sample [9/0]
H
0
N
S
P
Total
1-120
Ash
p
..
-
Stream HA Reference
52.89
4.1
43.40
1.17
0.58
<0.01
102.2
9.8
3.46
Stream F,i Reference
53.04
4.36
43.91
0.75
0.46
cO.01
102.5
8.9
0.98
Mooney Mooney NOM
6.3
The Mooney Mooney Dam NOM was also to be analysed by IHSS. However, the wet digestion
method whlch is used for HA and FA cannot be applied directly to NOM and is currently being
revised. The method to be developed will also analyse the ash composition.
4.7.5
XAD Fractionation
The XAD fractionation method is the classic concentration method for humic substances (see also
Chapter 2). The IHSS HA and FA samples were isolated using this method. This procedure was
therefore used to obtain humic substances from the Mooney Mooney NOM. The fractions were used
for NOM concentration and for experimental work.
A stock solution of about 4 g NOM in 500 mL water was prepared, resulting in a solution
concentration of 291. mgL-1 as DOC or a total mass of 145.5 mg organic carbon. The solution was
then desalted using an Amicon YC05 membrane (molecular weight cut-off 500 Da). According to
Amicon, this UF membrane does retain large salts such as phosphates and sulphates, but does not
retain a sipficant amount of smaller-sized salts. 310 mL of permeate were collected and discarded,
resulting in a loss of 5.0 mg organics (as DOC). Thus, 2.5% of organics, could be considered smaller
than the membrane pores.
The remaining solution volume of 190 mL was fractionated using the method of Leenheer (1981,
1996). Results are presented for the NOM sample in Figure 4.6.
Figure 4.6 Composition of
Moony Moony Dam NOM in
percent.
Copyright © 2001 by Andrea I. Schafer
102
MATERIALS AND METHODS
The sample has a high proportion of HA (47%) compared to fulvic and hydrophilic fractions (19%
each). T h s could account for the high microbiological activity in the Mooney Mooney Dam, w h c h
would result in a consumption of the more accessible fulvic and hydrophilic compounds. The relatively
high loss of organics in the XAD procedure is probably due to the presence of particulate organic
matter.
4.7.6
Cation Content of Organics
The cation content of the organic samples was determined using ICP-AES (see section 4.6.3 for
analytical details). Results are shown in Table 4.6.
The values per 100 mg DOC show the high salt content of NOM and its fractions. Whde the IHSS
samples and the XAD extracted HA and FA fractions of NOM are very low in cation content, the
NOM, the hydrophlic fraction of NOM, and the purified Aldrich HA have all very high cation
contents. The hydrophilc fraction has accumulated the entire salt content of the NOM sample. This
does not mean that all ions are associated with the hgdrophilic fraction, but due to the purification
method all ions remain in the hydrophilic sample. This needs to be considered when treatment data of
this sample are interpreted.
Table 4.6 Cation content of organics used The salt content ir per amount of DOC due to the stock rohtion
'
concentration. Vaher in bracketr are per l00 m g ~ .DOC, thus mg cationrper 100 mg DOC.
IHSS H A
IHSS FA
N O M HA
NOh.1 FA
A1 [mgL-l]
100
100
100
250.3
NOM
Aldrich 100
Hydrophhc
D O C [mgL-l]
NOM
kDa
22.1
12
114.5
0.10 (0.10)
0.02 (0.02)
0.58 (0.58)
0.24 (0.10)
0.07 (0.06)
0.47 (2.13)
0.28 (2.33)
Ca [mgL-l]
0.22 (0.22)
0 (0)
62.6 (62.6)
0.61 (0.24)
0.24 (0.21)
48.6 (219.9)
0.94 (7.83)
Fe [m&']
0.11 (0.11)
0 (0)
1.41 (1.41)
0.46 (0.18)
0.36 (0.31)
1.2 (5.43)
0.15 (1.25)
Na [mgL-'1
1.52 (1.52)
0.23 (0.23)
296 (296)
3.16 (1.26)
3.54 (3.09)
244 (1104.1)
12.3 (102.5)
I< [mgL-l]
0.55 (0.55) 0.41 (0.41)
52.4 (52.4)
2.16 (0.86)
1.19 (1.04)
1.43 (6.47)
0.47 (3.92)
4.7.7
High Performance Size Exclusion Chromatography (HPLC-SEC)
Size exclusion chromatography (SEC) enables the determination of the molecular size of organic
molecules. Samples were filtered through a 0.45 pm filter (Gelman Sciences Acrodiscs) prior to analysis.
The membrane filter material was Supor (Polyether-sulphone).
SEC was performed according to the method of Chin et al. (1994). A Shodex KW802.5 SEC column
PVaters Corp., Milford, MA., USA) was used and a Waters liquid chromatography system consisting of
the following components was used for the analysis: Waters 501 high pressure pump, Waters 717
autosampler, InterAction column temperature control oven, Waters 484 UV/VIS detector, and LVaters
Millenium 2.0 computer software package.
The mobile phase consisted of 200 mM phosphate at pH 6.8, adjusted to an ionic strength of 0.1 M
with high purity NaC1. The eluent was filtered through a preconditioned 0.22 pm membrane filter to
prevent interference from particulates. The system was operated at 1.0 mL/min and 30 ('C, with 200 pL
Copyright © 2001 by Andrea I. Schafer
103
Organics Characterisation
injections and detection at 260 nm. The mobile phase was degased for 30.minutes in an ultrasonic bath
prior to use.
The system was calibrated using polystyrene sulphonates (PSS) (Polysciences, NJ, USA). 1 gL-l
standards were prepared (35, 18, 8, 4.6 kDa). Blue Dextran, a high molecular weight polysaccharide
(approx. 2 000 kDa) and an acetone solution (1%) were used to determine the column's void volume
and total permeation volumes, respectively. The PSS's were detected at 224 nm (see Figure 4.7), the
acetone at 280 nm and the Blue Dextran at 260 nm. All samples were detected well inside the 15
min/sample run time.
Molecular Weight = 10
(Slope of L ~ n e RetentionTime)
+ Intercept of the Line.
(4.1)
The log of the molecular weight versus peak retention time for the PSS standards were plotted and
consistently yielded a straight line. By using the calibration equation:
The raw detector response versus retention time were converted to graphs of detector response versus
apparent molecular weight. The molecular weight determined for the organics used in this work is
shown in Figure 4.8. A number of observations can be made.
Surface water is the water from Mooney Mooney Dam prior to concentration and freeze drying,
whereas NOM is the redissolved powder of the same water. A. small, but nevertheless clear, increase in
molecular weight can be seen. It is thus obvious that the organic is being modified even using this
comparably "soft" concentration method.
Figure 4.7 HPLC-SEC PSS sta~zdardsin
single solzitions and as a mixttire.
100
1000
10000
Apparent Molecular Weight [Da]
The Aldrich HA has the largest size. Once this organic is purified by filtration through a 100 kDa
hIIVCO UF membrane, the size becomes comparable to the other organics. O f the purified
compounds, IHSS HA is the largest organic, and surface water the smallest. All organics have a size
distribution. The narrow peak at 300 Da is the salt peak. IHSS FA has a broader size distribution than
lHSS HA. Table 4.7 shows a summary of the peak molecular weight values. The values are the peak
height MW as determined from Figure 4.8.
Copyright © 2001 by Andrea I. Schafer