Bạn đang xem bản rút gọn của tài liệu. Xem và tải ngay bản đầy đủ của tài liệu tại đây (5.93 MB, 204 trang )
Sample Preparation Techniques
181
Method 5031 describes an azeotropic distillation technique for the determination
of nonpurgeable, water-soluble VOCs that are present in aqueous environmental
samples. The sample is distilled in an azeotropic distillation apparatus, followed by
direct aqueous injection of the distillate into a GC or GC-MS system. The method
is not amenable to automation. The distillation is time consuming and is limited to
a small number of samples.
Method 5032 describes a closed-system vacuum distillation technique for the
determination of VOCs that include nonpurgeable, water-soluble, volatile organics
in aqueous samples, solids, and oily waste. The sample is introduced into a sample
flask that is in turn attached to the vacuum distillation apparatus. The sample chamber
pressure is reduced and remains at approximately 10 torr (the vapor pressure of
water) as water is removed from the sample. The vapor is passed over a condenser
coil chilled to a temperature of 10°C or less. This results in the condensation of
water vapor. The uncondensed distillate is cryogenically trapped on a section of
1/8-in. stainless-steel tubing chilled to the temperature of liquid nitrogen (–196°C).
After an appropriate distillation period, the condensate contained in the cryogenic
trap is thermally desorbed and transferred to the GC or GC-MS using helium carrier
gas. The method very efficiently extracts organics from a variety of matrices. The
method requires a vacuum system along with cryogenic cooling and is not readily
automated.
Method 5035 describes a closed-system P&T for the determination of VOCs
that are purgeable from a solid matrix at 40°C. The method is amenable to soil/sediment and any solid waste sample of a consistency similar to soil. It differs from
the original soil method (Method 5030) in that a sample, usually 5 g, is placed into the
sample vial at the time of sampling along with a matrix-modifying solution. The sample
remains hermetically sealed from sampling through analysis as the closed-system
P&T device automatically adds standards and then performs the purge and trap. The
method is more accurate than Method 5030 because the sample container is never
opened. This minimizes the loss of VOCs through sample handling. However, it does
require a special P&T device. Oil wastes can also be examined using this method.
Method 5041 describes a method that is applicable to the analysis of sorbent
cartridges from a volatile organic sampling train. The sorbent cartridges are placed
in a thermal desorber that is in turn connected to a P&T device.
38. WHAT IS THE P&T TECHNIQUE?
The P&T method to isolate, recover, and quantitate VOCs in various environmental
sources of water has and continues to remain the premier technique for this class
of environmental contaminants. The technique was developed at the EPA in the early
1970s49 and remains the method of choice, particularly for environmental testing
labs that are regulatory driven. P&T has had the most success with drinking water
samples when combined with gas chromatography and element-specific detectors.
GC detectors will be discussed in Chapter 4. In this section, we will discuss the
EPA methods that use the P&T technique to achieve the goals of TEQA. EPA Method
502.2 summarizes the method as follows:50
© 2006 by Taylor & Francis Group, LLC
182
Trace Environmental Quantitative Analysis, Second Edition
Highly volatile organic compounds with low water solubility are extracted (purged)
from the sample matrix by bubbling an inert gas through a 5 mL aqueous sample.
Purged sample components are trapped in a tube containing suitable sorbent materials.
When purging is complete, the sorbent tube is heated and back-flushed with helium to
thermally desorb trapped sample components onto a capillary gas chromatography
(GC) column. The column is temperature programmed to separate the method analytes
which are then detected with a photoionization detector (PID) and an electrolytic
conductivity detector (ElCD) placed in series. Analytes are quantitated by procedural
standard calibration.… Identifications are made by comparison of the retention times
of unknown peaks to the retention times of standards analyzed under the same conditions used for samples. Additional confirmatory information can be gained by comparing the relative response from the two detectors. For absolute confirmation, a gas
chromatography/mass spectrometry (GC/MS) determination according to EPA Method
524.2 is recommended.
The classical purge vessel is shown in Figure 3.10. Usually, 5 mL of an environmental aqueous sample is placed in the vessel. The sample inlet utilizes a twoway valve. A liquid-handling syringe that can deliver at least 5 mL of sample is
connected to this sample inlet, and the sample is transferred to the purge vessel in
a way that minimizes the sample’s exposure to the atmosphere. Note that the
incoming inert purge gas is passed through a molecular sieve to remove moisture.
A hydrocarbon trap can also be inserted prior to the purge vessel to remove traces
of organic impurities as well. The vessel contains a fritted gas sparge tube that serves
to finely divide and disperse the incoming purge gas. These inert gas bubbles from
the purging provide numerous opportunities for dissolved organic solutes to escape
to the gas phase. The acceptable dimensions for the purge device are such, according
to EPA Method 502.2, that it must accommodate a 5-mL sample with a water column
at least 5 cm deep. The headspace above the sample must be kept to a minimum of
<15 mL to eliminate dead-volume effects. The glass frit should be installed at the
base of the sample chamber, with dispersed bubbles having a diameter of <3 mm
at the surface of the frit.
Figure 3.10 also depicts a typical trap to be used in conjunction with the purge
vessel. The trap must be at least 25 cm long and have an inside diameter of at least
0.105 in., according to Method 502.2. The trap must contain the following amounts
of adsorbents:
•
•
•
One third of the trap is to be filled with 2,6-diphenylene oxide polymer,
commonly called Tenax.
One third of the trap is to be filled with silica gel.
One third of the trap is to be filled with coconut charcoal.
It is recommended that 1 cm of methyl silicone-coated packing be inserted at
the inlet to extend the life of the trap. Method 5030B from the SW-846 series is
more specific than Method 502.2 and recommends a 3% OV-1 on Chromosorb-W,
60/80 mesh, or equivalent. Analysts who do not need to quantitate dichlorodifluoromethane do not need to use the charcoal and can replace this charcoal with more
Tenax. If only analytes whose boiling points are above 35°C are to be determined,
© 2006 by Taylor & Francis Group, LLC
Sample Preparation Techniques
Optional
foam
trap
183
Exit 1/4 IN.
O.D.
14 mm O.D.
Inlet 1/4 IN.
O.D.
Sample inlet
2-Way syringe valve
17 cm. 20 gauge syringe needle
1/4 IN.
O.D. Exit
6 mm. O.D. rubber septum
10 mm. O.D.
10 mm. 14 mm. O.D.
Inlet
1/4 IN.
O.D.
1/16 IN. O.D.
Stainless steel
13% Molecular
sieve purge
gas filter
Purge gas
flow
control
10 mm. Glass frit
medium porosity
Packing procedure
Glass 5 mm
wool
Activated 7.7 cm
charcoal
7 Ω/foot
Resistance
wire wrapped
solid
(double layer)
Grade 15 7.7 cm
silica gel
Tenax 7.7 cm
3% 0 V–1 1 cm
Glass wool 5 mm
Trap inlet
15 cm
7 Ω/foot
Resistance
wire wrapped
solid
(single layer)
8 cm
Construction
Compression
fitting nut
and ferrules
Thermocouple/
controller
sensor
Electronic
temperature
control and
pyrometer
Tubing 25 cm
0.105 IN. I.D.
0.125 IN. O.D.
Stainless steel
FIGURE 3.10 Schematic of a purge-tube-and-trap configuration.
both the charcoal and the silica gel can be eliminated and replaced with Tenax. The
trap needs to be conditioned at 180°C prior to use and vented to the atmosphere
© 2006 by Taylor & Francis Group, LLC
184
Trace Environmental Quantitative Analysis, Second Edition
Helium
N2 liq .
ECD
Trap
Gas chromatograph
Sample
Sparger
Helium
Vent
(a)
N2 liq .
ECD
Trap
Gas chromatograph
Heat
Helium
Vent
(b)
FIGURE 3.11 Schematics of the purge, trap, and thermal desorption for trace VOCs in water.
instead of the analytical GC column. It is also recommended that the trap be
reconditioned on a daily basis at the same temperature. Tenax is a unique polymer
and offers the advantage that water is not trapped to any great extent.
Commercially available P&T units are fully automated and consist of a bank of
purge vessels with switching valves that enable one vessel to be purged after another.
Figure 3.11 is a schematic diagram that clearly depicts the purge, trap, and thermal
desorption steps involved in this technique. A six-port valve, placed after the trap
and interfaced to the injection port of a GC, provides the needed connection. Referring to Figure 3.11a, the P&T step is shown. Inert gas, usually helium, enters the
© 2006 by Taylor & Francis Group, LLC
Sample Preparation Techniques
185
purge vessel while the trap outlet is vented to the atmosphere. Meanwhile, GC carrier
gas flows through one side of the six-port valve directly to the GC. The purge vessel
and the trap are generally kept at ambient temperature. Some commercially available
units provide for a heated purge vessel. Referring to Figure 3.11b, the direction of
gas flow during the desorb step is illustrated. The valve is turned and inert gas enters
and passes over the trap in a direction opposite to the P&T step. The trap is rapidly
heated to the required final temperature and the trap outlet is directed to the injection
port of the GC. GCs that are equipped with cryogenic cooling can deposit the VOCs
from the trap to the GC column inlet as a plug. The GC can then be temperature
programmed from the cryogenic temperature to the final temperature, and hence
complete the gas chromatographic separation of all VOCs.
Operating conditions drawn from EPA Method 5030B for the P&T system differ
slightly based on which determinative method is used and are presented in the
following:
Analysis Method
Purge gas
Purge gas flow rate (mL/min)
Purge time (min)
Purge temperature (°C)
Desorb temperature (°C)
Back-flush inert gas (mL/min)
Desorb time (min)
8015
8021 or 8260
N2 or He
20
15.0
85
180
20–60
1.5
N2 or He
40
11.0
Ambient
180
20–60
4
39. HOW DOES ONE GO ABOUT CONDUCTING
THE P&T TECHNIQUE?
Let us assume that the decisions of whether to screen the sample prior to conducting
P&T have been made and that we are ready to actually perform the technique. We
have available in our laboratory organic-free water, the determinative method has
been defined, and we have reviewed EPA Method 5000 for guidance with respect
to internal and surrogate standards. Environmental samples such as groundwater,
drinking water, wastewater, and so forth, have already been collected, stored in
capped bottles with minimum headspace, free of solvent fumes, and stored at 4°C.
If the sample was found to be improperly sealed, it should be discarded. All samples
should be analyzed within 14 days of receipt at the laboratory. Samples not analyzed
within this period must be noted and data are considered to represent a minimum
value. This is the essence of the so-called holding times. Holdings times, like
detection limits that were discussed in Chapter 2, are a controversial topic between
the regulatory agency and contracting laboratory. Using Method 5030B for guidance,
the sequence of steps needed to carry out the method are as follows:
•
Initial calibration: The P&T apparatus is conditioned overnight with an
inert gas flow rate of at least 20 mL/min and the Tenax trap held at a
© 2006 by Taylor & Francis Group, LLC
186
Trace Environmental Quantitative Analysis, Second Edition
•
•
•
•
temperature of 180°C. The P&T apparatus is connected to the GC, and
each purge vessel is filled to its total volume with organic-free water.
Methanolic solutions containing VOCs at high enough concentration such
that only microliter aliquots need to be added to the water in each purge
vessel are added to a series of purge vessels. A blank and set of working
calibration standards are prepared right in the purge vessels. Internal
standards and surrogates are also added as their methanolic solutions as
standard references in a manner similar to that of the analytes to be
calibrated. Refer to Chapter 2 for a discussion of the various modes of
instrumental calibration.
Conduct the P&T for the set of blanks and working calibration standards.
Initial calibration verification (ICV): Prepare one or more purge vessels
that contain the ICV. Sometimes, it is advantageous to prepare ICVs in
triplicate to enable a preliminary evaluation of the precision and accuracy
of the calibration to be made. The ICV criteria are usually given in a
determinative method such as in the 8000 series of SW-846. These criteria
must be met before real samples can be run.
Adjust the purge gas flow rate referring to the guidance given in the above
table.
Sample delivery to the purge vessel: Remove the plunger from a 5-mL
liquid-handling syringe and attach a closed syringe valve. Open the sample
or standard bottle, which has been allowed to come to ambient temperature. Carefully pour the sample into the syringe barrel to just short of
overflowing. Replace the syringe plunger and compress the sample. Open
the syringe valve and vent any residual air while adjusting the sample
volume to 5.0 mL. This process of taking an aliquot destroys the validity
of the liquid sample for future analysis. If there is only one sample vial,
the analyst should fill a second syringe at this time to protect against
possible loss of sample integrity. Alternatively, carefully transfer the
remaining sample into a 20-mL vial and seal with zero headspace. The
second sample is maintained only until such time when the analyst has
determined that the first sample has been analyzed properly. In this way,
the VOC content of the environmental sample is preserved during the
transfer to the purge vessel.
A GC-gram from EPA Method 502.2 that depicts the separated peaks from both
the PIDs and ElCDs is shown in Figure 3.12. Both chromatograms are overlayed
and show the different response for each chromatographically resolved peak between
detectors.
40. DO ALL VOCs PURGE WITH THE SAME RATE?
No, they do not, and until recently, little was known about the kinetics of purging
for priority pollutant VOCs using conventional P&T techniques. We describe the
findings of Lin et al.,51 who demonstrated that first-order kinetics are followed for
the removal of VOCs from P&T vessels. The authors make the point that because
© 2006 by Taylor & Francis Group, LLC
10.0
Time, min
30.0
20.0
18
43
+
52
35 40.0 44
47
31
41
33 36
45 49
37 46 48
+
34
24
50.0
55.0
57
PID
26
+
27
20
8
13
12 15 19
16
9
ELCD
10
14
7
2
3
1
5
4
17
11
32
50
Sample Preparation Techniques
0.0
51
53
56
21
22
23
25
30
29
28
55
42
58
40 44
45
38
39
54
FIGURE 3.12 Capillary gas chromatogram for the separation of VOCs using EPA Method 502.2.
187
© 2006 by Taylor & Francis Group, LLC
188
Trace Environmental Quantitative Analysis, Second Edition
EPA methods such as 524.2 and 624 require the use of internal standards to obtain
relative response factors, the percent purge efficiency is never considered. Some 28
priority pollutant VOCs were studied. Each VOC, dissolved in methanol, was spiked
into high-purity water at concentration levels of 1 and 10 ppb, with a 10 ppb internal
standard. Spiked samples were purged for 11 min. A GC-MS was used, and the
absolute peak area for the characteristic primary ion was used to quantitate. Experimentally, samples were purged with helium for 11 min and GC-MS data were
obtained. These purged samples were then purged a second time for the same 11-min
duration under identical conditions. Peak areas were obtained in the same manner
as for the first purge.
Let us begin to construct a mathematical view of the kinetics of P&T by first
defining the purge ratio, Pi, as being the ratio of the mean GC-MS area count for
the first purge to the mean GC-MS area count for the second purge according to
A /5
A (ave) ∑
P=
=
A (ave )
∑ A /5
5
i
1
i
2
i
j
5
j
1
ij
2
ij
Each peak area for the ith analyte was obtained by averaging peak areas over j
replicate purges, with j taken from 1 to 5. A selected list of VOCs, along with each
VOC retention time, purge ratio, and coefficient of variation (refer to Chapter 2)
from their work, is given in the following table:
VOC
Chloromethane
1,2-Dichloroethane
1,1,1,2-Tetrachloroethane
1,2,4-Trichlorobenzene
Acetone
2-Butanone
tR (min)a
Pi
% RSD
1.82
9.99
16.98
26.30
4.30
8.34
339
2.21
6.06
3.64
1.05
1.03
36
6
9
8
2
5
a30 m × 0.53 mm DB-624 (J&W Scientific); refer to EPA
Method 524.2 for GC column temperature program and other
GC conditions.
The low percent RSDs, except for the very volatile chloromethane, indicate that
Pi is constant over the range of concentrations studied. According to these authors,
consistent values of Pi would indicate that the purging VOCs from an aqueous
solution can be mathematically described by first-order chemical kinetics. How fast
any of the 28 VOCs studies are removed from a fixed volume of an aqueous sample
such as groundwater in a conventional purge vessel is expressed as –dC/dt, where
the negative sign indicates that the concentration of a VOC decreases with time.
First-order kinetics suggests that the rate is directly proportional to the concentration
of each VOC.52 Expressed mathematically,
© 2006 by Taylor & Francis Group, LLC
Sample Preparation Techniques
189
dC
= − kC
dt
(3.32)
where C is the analyte concentration at any time t and k is the first-order rate constant.
Differences in the magnitude of k depend on the chemical nature of the particular
VOC and the degree of intermolecular interaction with the sample matrix — in this
case, a relatively clean water sample. Equation (3.32) is first rearranged as
dC
= − k dt
C
Integrating between the initial concentration C1 and the concentration after the
first purge, C2, on the left side, while integrating between time t = t0 and time t =
t0 + ∆t on the right side is shown below. Identical integration is also appropriate
between C2 and C3. ∆t is the time it takes to purge the sample. We then have
∫
C2
C1
dC
= −k
C
∫
t0 + ∆t
dt
t0
Evaluating the definite integrals for both sides of the equation yields
ln(C 2 − C1 ) = − k ∆t
Utilizing a property of logarithms gives
ln
C2
= − k ∆t
C1
Expressed in terms of exponents,
C2
= e − k ∆t
C1
∴C2 = C1e
(3.33)
− k ∆t
Equation (3.33) implies that C2 can be expressed in terms of C1, k, and ∆t.
The experiment starts with each VOC at a concentration C1. After purge 1, each
analyte is at concentration C2. After purge 2, each analyte is at concentration C3.
This is depicted by
C1
Purge 1→ C 2
© 2006 by Taylor & Francis Group, LLC
C2
Purge 2 → C 3
190
Trace Environmental Quantitative Analysis, Second Edition
Because VOCs are continuously removed from a fixed volume of aqueous
sample, it is also true that
C 3 < C 2 < C1
The peak area obtained after the first purge A1 is proportional to the difference
in concentration between C1 and C2; that is,
A1 ∝ C1 − C2
Substituting for C2 using Equation (3.33) gives
A1 ∝ C1 (1 − e − k ∆t )
Likewise, the peak area obtained for each analyte after the second purge is
proportional to the difference in concentration between C2 and C3 according to
A2 ∝ C2 − C3
Substituting for C3 using Equation (3.33) gives
A2 ∝ C2 (1 − e − k ∆t )
The ratio of peak areas between the first and second purges for a given analyte,
P, can be related to these differences in concentration as follows:
P=
C1 − C 2 C1
=
C2 − C3 C2
Substituting for Equation (3.33) yields
P=
C1
C1e − k ∆t
This simplifies to
P = e k ∆t
(3.34)
Equation (3.34) suggests that P is always greater than 1 and is independent of
the initial analyte concentration. The development of Equation (3.34) has assumed
first-order kinetics. The authors observed good precision over the five replicate
experiments for all 28 VOCs studied, with the exception of the very volatile chloromethane.
If P is determined experimentally, by first obtaining the GC-MS peak area, A1,
and then obtaining A2, Equation (3.34) can be solved for the first-order rate constant, k.
© 2006 by Taylor & Francis Group, LLC
Sample Preparation Techniques
191
1
In addition, when C 2 = 2 C1, the time it takes for half of the analyte to be
removed can be found according to
1/ 2
C1 C1
C
= 1 C1 = 2 1 = e kt
C2
C1
2
1/ 2
2 = e kt ]
(3.35)
ln 2 = kt1/ 2
∴ t1/ 2 =
1
ln 2
k
Using Equation (3.35), the t1/2 values given for the three VOCs from the author’s
work are shown in the following table:
VOC
Chloromethane
1,1,1,2-Tetyrachloroethane
Acetone
t1/2 (min)
1.3
4.2
156
A purge time, ∆t = 11 min, is used in indicated EPA methods. For hydrophobic
VOCs, the t1/2 seems to be too low, and for hydrophilic VOCs, the t1/2 seems to be
too high. Other generalizations emerge with respect to the chemical nature of the
VOCs studied. The effect of deuterating 1,2-dichlorobenzene yields values for t1/2
that are quite close to each other, 5.6 and 5.8 (1,2-dichlorobenzene-d4), respectively.
Differences in structural isomers such as 1,1,2,2- vs. 1,1,1,2-tetrachloroethanes are
reflected not only in their volatility in water, but also in their gas-phase stability of
both neutral and ionic states. It would also appear that the choice of internal standard
should reflect the kinetics of purging in addition to GC relative retention time.
Choosing internal standards based on relative retention times has been the hallmark
of most EPA methods. This work would suggest that t1/2 also be taken into consideration when deciding on which internal standard to use when using P&T techniques.
We leave sample prep for VOCs and return to SVOCs. Conventional sample
extract cleanup techniques are introduced followed by SFE and then SPE. We already
discussed how LLE is used to remove unwanted neutral interferences when the target
analyte of interest is a weak acid and can therefore be ionized. Cleanup introduced
here refers to the need to remove neutral polar interferences from neutral targeted
SVOC analytes.
41. WHAT IS CLEANUP?
It is not the use of detergent to remove dirt. It is the removal of chemical interferences
from the extract following any of the extraction techniques discussed earlier. The
© 2006 by Taylor & Francis Group, LLC