Bạn đang xem bản rút gọn của tài liệu. Xem và tải ngay bản đầy đủ của tài liệu tại đây (5.93 MB, 204 trang )
Sample Preparation Techniques
243
with an Saq of between 7.5 and 10 ppm. Endrin is a member of the cyclodiene
insecticides, possessing the characteristic endomethylene bridge structure. The Saq
for endrin is 0.23 ppm, which is similar to that of aldrin, dieldrin, and heptachlor
epoxide, but greater than that of heptachlor and chlordane. Endrin is known to
partially break down to endrin aldehyde and endrin ketone in packed GC columns.
It is not much of a problem when capillary columns are used. Methoxychlor, a
substituted diphenyl trichloroethane similar to DDT, has a Saq of between 0.1 and
0.25 ppm, and it is about 100 times more soluble in water than DDT. Using ethyl
acetate as an elution solvent and some of the early chemically bonded silica, of 60 Å
pore size and 40-µm irregular particle size (Separalyte, Analytichem International),
this author conducted a series of systematic studies of lindane, endrin, and methoxychlor (L, E, M). The work was focused on a sample matrix consisting of distilled
deionized water (DDI), and the analysis was performed on a gas chromatograph that
used a packed column and electron-capture detector.
One such study is reported here. Nine cartridges were packed with an octylbonded silica, C8 Separalyte. The mass of the sorbent packed in each cartridge varied
from a low of 20 mg to a high of 320 mg. In each cartridge, after methanol
conditioning, a spiked aqueous sample was passed through the packed bed under a
reduced-pressure SPE manifold. Then 8.8 µL of a methanolic reference standard
containing L, E, and M, each at a concentration of 100 ng/µL (ppm), was added to
approximately 60 mL of DDI. This spiking gave a concentration level in the aqueous
sample of 15 pg/µL (ppb). This concentration level is considered low enough to
approximate the realm of TEQA. After the sample was passed through the nine
replicate RP-SPE cartridges, two 500-µL aliquots of ethyl acetate were used to elute
the sorbed analytes into a 1-mL receiver volumetric flask. The contents of the
volumetric flask were quantitatively transferred to a 2-mL GC autosampler glass
vial, and 2 µL of this eluent was injected into a GC that contained a packed
chromatographic column. Only one injection per GC vial was made in this particular
study, so that variation in the percent recovery reflects only the random error associated with the RP-SPE process only. Table 3.12 lists the percent recovery for all
three OCs from each of the nine cartridges. The concentration of analyte that
constituted a 100% recovery was calculated and not actually measured; hence, there
is no contribution to the relative standard deviation from the control standard, as
shown in Equations (2.30) and (2.31). It is evident from review of the percent
recovery results shown in Table 3.12 that breakthrough was not reached even in the
case where only 20 mg of sorbent was taken to prepare the packed cartridge.
The mean percent recovery can be found from these nine replicate SPEs. The
standard deviation in the mean percent recovery is then found using the fundamental
equation for calculating standard deviations. If replicate injections per GC vial were
made, a pooled standard deviation relationship, as given in Equation (2.33), would be
most appropriate. From the standard deviation, a relative standard deviation or coefficient of variation is obtained, followed by calculation of the corresponding confidence
interval. For the percent recoveries shown in Table 3.12, the following statistical evaluation was obtained:
© 2006 by Taylor & Francis Group, LLC
244
Trace Environmental Quantitative Analysis, Second Edition
Mean
% Recovery
OC
Lindane
Endrin
Methoxychlor
RSD
(%)
Confidence Interval
(95%)
96.8
107.3
94.1
8.3
14.3
8.2
6.2
11.8
5.9
TABLE 3.12
Percent Recoveries of OCs Using a C8-Bonded
Silica Sorbent from Spiked Water
Sorbent (mg)
Lindane
Endrin
Methoxychlor
115
291
225
241
98
20
52
130
320
94.3
97.1
106
100
100
100
77.1
97.1
100
111
87.3
133
111
122
106
87.3
113
95.2
93.2
80.8
107
100
100
93.2
86.3
93.2
93.2
Note: 8.8 µL of 100 ppm each of lindane, endrin, and methoxychlor, dissolved in methanol, was added to approximately
60 mL of distilled deionized water. The spiked samples were
passed through the cartridge, previously conditioned with
methanol, and eluted with two 500-µL aliquots of ethyl acetate.
The volume of eluent was adjusted to 1.0 mL, and 2 µL of
eluent was injected into a GC.
The mean percent recoveries in this replicate series of RP-SPEs are very high
and represent a statement of accuracy in the measurement. A relative standard
deviation of 8 or 14% among replicate SPEs represents the precision of the method.
The confidence interval states that of the next 100 SPEs, 95 of these should fall
within the interval specified. For example, if 100 additional percent recoveries using
the SPE method could be performed for lindane, one could expect that 95 would
fall within 96.8 ± 6.2%.
This high and reproducible percent recovery for the isolation and recovery of
lindane from water strongly suggests that RP-SPE is very appropriate as a sample
preparation method for this analyte. Lindane is of continued interest to environmental
and toxicological scientists. One such study, discussed next, taken from the author’s
work, involves the isolation and recovery of lindane from homogenized myometrial
tissue suspended in an aqueous matrix.
© 2006 by Taylor & Francis Group, LLC
Sample Preparation Techniques
245
68. WAS LINDANE ISOLATED AND RECOVERED FROM A
BIOLOGICAL MATRIX USING RP-SPE? IF SO, HOW?
One hundred microliters of a methanolic solution containing 5 ng/µL (ppm) lindane
was placed in a 1-mL volumetric flask half filled with iso-octane, while 100 µL of
the same lindane reference standard was added to approximately 70 mL of distilled
deionized water (DDI). One microliter of the former solution containing 500 ng of
lindane was injected into a GC, and the resulting peak area served to define a control
that represents a 100% recovery of lindane. One microliter of the 1-mL eluent from
performing the RP-SPE of the spiked DDI was also injected into the same instrument,
and the resulting peak area served to define the spiked recovery sample. In this study,
quadruplicate injections of the control yielded a mean concentration of 400 ppb
lindane from interpolation of the least squares calibration curve. Seven replicates of
the eluent from the spiked DDI sample were injected, and a mean concentration of
406 ppb lindane was obtained from interpolation of the same calibration curve.
Equation (2.57) was used to find a 106% recovery of lindane. RSDs in the % recovery
were found by propagating random error between spiked samples and control reference standards. A complete result can be given by stating both the accuracy, 106%
recovery, and the relative standard deviation of 15.6%. Myometrium samples were
then prepared and lindane appeared as expected. This work showed that lindane
could easily be isolated and recovered from an aqueous matrix and confirmed the
earlier work on lindane isolation and recovery, discussed previously. One would be
led to believe that this is a robust sample preparation method for lindane because it
can be reproduced with confidence. Thus, a subsequent request for additional sample
analyses does not require extensive QC and should merely report the analytical
results.
69. WHAT DOES A SAMPLE ANALYSIS REPORT LOOK LIKE?
In addition to a tabular format for the determination of lindane in each of the
myometrial tissue samples submitted, a method summary should be included. A
one-paragraph method summary serves to inform the reader as to how the samples
were handled once they arrived in this analyst’s laboratory. The summary should
also provide a brief overview of the sample preparation and a brief description of
the determination technique used. The following report illustrates these concepts.
REPORT ON THE QUANTITATIVE DETERMINATION
TISSUE SUSPENDED CELLS IN SALINE
OF
LINDANE
IN
MYOMETRIAL
Summary of Method
The entire contents (1 mL) of sample that was received by the client were refrigerated
upon receipt until the sample was prepared for analysis. Upon thawing, the entire
contents of each sample were added to a reservoir that contained approximately 70 mL
of distilled deionized water. This aqueous solution was passed across a previously
conditioned octadecyl-bonded silica sorbent (C18RPSiO2). The retained analyte was
© 2006 by Taylor & Francis Group, LLC
246
Trace Environmental Quantitative Analysis, Second Edition
eluted off of the sorbent with two 500-µL aliquots of pesticides-residue-grade isooctane. The iso-octane eluent was then passed through a second SPE cartridge. This
second cartridge was packed with approximately 0.5 g of anhydrous sodium sulfate.
The volume of eluent that now contained the recovered analyte was adjusted to a final
volume of 1.0 mL using a volumetric flask. Three microliters of this eluent was injected
via autosampler into an Autosystem Gas Chromatograph (PerkinElmer) incorporating
an electron-capture detector. A 30 m × 0.32 mm capillary GC column containing DB-5
(J&W Scientific) was used to separate the organics in the eluent. This instrument is
abbreviated C-GC-ECD to distinguish it from other gas chromatographs in our laboratory. The column was temperature programmed following injection from 200 to
270°C at a rate of 10°C/min. The C-GC-ECD is connected to a 600 Link (PE-Nelson)
interface module. This interface, in turn, is connected to a 386 personal computer. This
PC uses Turbochrom® (PE-Nelson) software for data acquisition, processing, and
control. A method specific for lindane was written, and one peak was identified within
a 3% relative time interval (retention time window). A retention time, t[R], of 1.8 min
was consistently reproduced using autosampler injection.
Calibration
A series of calibration or working standards were prepared from the methanolic stock
solution containing lindane. This stock solution was prepared by carefully weighing
out pure solid lindane on an analytical balance. These standards were injected into the
C-GC-ECD from lowest to highest lindane concentration. The peak at 1.8 min was
identified as a reference peak in the Turbochrom software, and a narrow t[R] window
was defined around the t[R] of the apex of the peak. A calibration curve was constructed
using a least squares regression algorithm in the software. A correlation coefficient of
0.9990 was obtained.
Sample Analysis
The sequence within which the samples were run after calibration was created by using
the Sequence File Editor within Turbochrom. Samples are injected via autosampler
according to the instructions in the Sequence File. After the sequence is completed, a
Summary File is created and the analytical results are reported within this summary
format. In the Summary File, for each sample analyzed, the following is given:
(a) The sample number
(b) The concentration of lindane in ppb for 1.0 mL of eluent, interpolated from the
external standard mode of instrument calibration
(c) The retention time for lindane, t[R], in minutes
(d) The integrated peak area, in microvolts-seconds
The reported concentration is that for each eluent and should be multiplied by the
eluent volume (in this case, 1.0 mL) to obtain the number of nanograms of lindane
found. The number of nanograms found divided by the volume of the aqueous sample
used in RP-SPE gives the reported concentration in ppb for lindane. The reported
concentration should be divided by the volume of the sample to obtain the reported
concentration in ppb for lindane. The reported concentration should also be divided
by the percent recovery, expressed as a decimal, in order to find the true and final
concentration of lindane in the original sample.
© 2006 by Taylor & Francis Group, LLC